Drawing: | RCT001/5 |
EOS Drawing: | SCH-4025-1 |
Manual: | MM page 10-5 |
Scale: | 2000 counts per millimeter |
Motor: | |
Motor Encoder: | |
Limits: | two level on each side |
Controller: | |
X Stage: | ATS100-100 Aerotech |
X Drive Amplifier: | Amp 1 |
Homing: | 2mm/rev lead screw, 1000 line encoder For the ATS-100 |
Range: | -202500 to +9500 (106mm) |
MaxSpeed: | 10000 cts/sec) |
Encoder Zero: | closed to positive end probe out of field of view |
Note: X-axis, positive motion is in negative declination.
Homing may take a minute.
Scale: | 2000 counts per millimeter |
Motor: | |
Motor Encoder: | |
Overall Encoder: | |
Limits: | none |
Controller: | |
Y Stage: | ATS100-200 |
Y Drive Amplifier: | Amp 2 |
Homing: | 2mm/rev lead screw, 1000 line encoder for the ATS-100 |
Range: | -8500 to +403000 range is 205.75mm |
Max Speed: | 10,000 cts/sec |
EncoderZero: | close to negative end (out of FOV) |
Probe is centered in the FOV at X = -189350, Y = +199500
Scale: | 8000 counts per millimeter |
Motor Encoder: | |
Limits: | |
Controller: | |
Z Stage: | ATS03005 |
Z Drive Amplifier: | Amp_3 |
Homing: | 0.5mm/rev lead screw, 1000 line encoder for the ATS-300 |
Range: | -8500 to +429500, range : 54.75mm |
MaxSpeed: | 15000 cts/sec |
EncoderZero: | near negative range of motion |
Guidercollimated with camera: | -8000? |
Note: May take 2 minutes to home.
EOS 11Sep2003 ProbeX -3350 X counts = +20 arc-seconds dec ProbeY -3300 Y counts = -20 arc-seconds HA ProbeZ +120000 Y counts = +.325 mm Focus
See Sections 5.7.8 5.7.9 and 5.1.6 for operating software.